Repeatability and reproducibility of Galilei measurements in normal keratoconic and postrefractive corneas
Küçük Resim Yok
Tarih
2014
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Elsevier Science Bv
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
Objective: To assess the repeatability and reproducibility of the anterior segment measurements performed with a dual Scheimpflug analyzer (Galilei) in normal, keratoconic and post-refractive surgery corneas. Methods: To evaluate the repeatability, two additional measurements were performed by the first examiner. To assess reproducibility, this was later followed by a single reading by the second examiner. The following parameters were recorded and evaluated in this study; central corneal thickness (CCT), thinnest corneal thickness (TCT), mean total corneal power (TCP) in central (0-4 mm), mean posterior corneal power (PCP) in central (0.5-2 mm), anterior and posterior elevation (best fit sphere [BFS]) in central 8 mm anterior and posterior eccentricity (epsilon(2)) in central 8 mm. Repeatability and reproducibility for each corneal parameter was assessed using the Bland-Altman analysis. Results: Each of the three groups was consisted of 20 subjects (totally 60 patients, 30 men and 30 women). The 95% LoA for repeatability was very small, indicating small discrepancies between measurements related to CCT. Acceptable repeatability was also achieved for the other parameters in each group. However, the 95% LoA for value TCP was larger in keratoconic eyes. The 95% LoA for reproducibility was also very small, and acceptable for all measured parameters in each group. In addition, the 95% LoA was larger for the measurement of CCT and TCT for postrefractive corneas. Conclusions: The anterior segment measurements provided by Galilei showed good repeatability and reproducibility for normal, keratoconic and postrefractive corneas. (C) 2014 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.
Açıklama
Anahtar Kelimeler
Galilei; Keratoconus; Repeatability; Reproducibility
Kaynak
Contact Lens & Anterior Eye
WoS Q Değeri
Q3
Scopus Q Değeri
Q1
Cilt
37
Sayı
5