Repeatability and Reproducibility of Anterior Segment Measurements in Normal Eyes Using Dual Scheimpflug Analyzer

Küçük Resim Yok

Tarih

2015

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Turkish Ophthalmological Soc

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Objectives: To assess the repeatability and reproducibility of anterior segment measurements including aberrometric measurements provided by a dual Scheimpflug analyzer (Galilei) system in normal eyes. Materials and Methods: Three repeated consecutive measurements were taken by two independent examiners. The following were evaluated: total corneal power and posterior corneal power, corneal higher-order wavefront aberrations (6.0 mm pupil), pachymetry at the central, paracentral, and peripheral zones, and anterior chamber depth (ACD). Repeatability was assessed by calculating the within subject standard deviation, precision, repeatability, and intraclass correlation coefficient (ICC). Bland-Altman analysis was used for assessing reproducibility. Results: Thirty eyes of 30 patients were included. The best ICC values were for corneal pachymetry and ACD. For both observers, acceptable ICC was also achieved for the other parameters, the only exceptions being posterior corneal astigmatism and total high order aberration. The 95% LoA (Limits of Agreement) values for all measurements showed small variability between the two examiners. Conclusion: The Galilei system provided reliable measurements of anterior segment parameters. Therefore, the instrument can be confidently used for routine clinical use and research purposes.

Açıklama

Anahtar Kelimeler

Galilei; intraobserver; interobserver repeatability; wavefront aberrations

Kaynak

Turk Oftalmoloji Dergisi-Turkish Journal Of Ophthalmology

WoS Q Değeri

N/A

Scopus Q Değeri

N/A

Cilt

45

Sayı

6

Künye